Quality Assurance and Quality Control

Instrumentation for stable isotope analysis has expanded to include both the original Finnigan Delta-S and now two GV Instruments IsoPrime isotope ratio mass spectrometer. With the addition of the GVI instruments and associated peripherals, we have automated stable isotope analysis of most organic (and some inorganic) samples for carbon-13 and nitrogen-15. The workhorse of continuous flow measurements of solid samples for carbon and nitrogen isotopes is done by the GVI IsoPrime and a Elementar Cube elemental analyzer, combined with a diluter and reference gas box.

OLYMPUS DIGITAL CAMERA

Samples for automated isotope analysis are first weighed out into tin boats to the nearest 0.001 mg on a Kahn C33 or a Sartorious micro electronic balance. During a sequence run by the mass spectrometer, each sample is flash combusted at 1800ÂșC in the Elementar CN analyzer; the combustion gases (CO2, N2 and H2O) are trapped and separated, and introduced into the mass spectrometer, with water removed in a chemical trap. The gases of interest are then introduced into the mass spectrometer for isotope analysis and the rest pumped away. The sample isotope ratio is compared to a secondary gas standard, whose isotope ratio has been calibrated to international standards. For 13CV-PDB the gas was calibrated against NBS 20 (Solenhofen Limestone), NBS 21 (Spectrographic Graphite), and NBS 22 (Hydrocarbon Oil); for 15Nair the gas was calibrated against atmospheric N2 and IAEA standards N-1, N-2, and N-3 (all are ammonium sulfate standards). All international standards were obtained from the National Bureau of Standards in Gaithersburg, MD. A two point calibration is also done using USGS 40 and USGS 41 standards (Copeland laboratory

International standards used for water samples include V-SMOW, GISP and SLAP, which were utilized to calibrate the secondary gas standard. In the past, water samples were prepared using the guanidine hydrochloride technique for oxygen-18 and the zinc reduction technique for deuterium (zinc obtained from John Hayes, Indiana University). With the addition of the GVI MultiFlow and ChromeHD systems, we are now able to automate both procedures. Oxygen-18 analysis is done via CO2 equilibration and deuterium analysis is done via pyrolysis in the ChromeHD system.

When running gas samples on the Finnigan Delta-S, as a daily check on instrument performance we run a second gas (lecture N2 or CO2) that is isotopically distinct from our standard gas. If the isotope values are within 0.05 per mil of its long-term record, analysis proceeds; otherwise, further analysis stops until any problems are resolved. Internal precision for the instrument is +0.014 per mil. For solid continuous flow samples, a suite of in-house standards are first analyzed. If they fall within laboratory specifications, client sample analysis then proceeds.

Required external precision of a sample (i.e. replicate analysis) for either 15N or 13C is 0.2 per mil. Typically our precision is better than 0.1 per mil for well-ground organic tissue samples using the trapping box.

Samples run in continuous flow mode are currently within 0.2 per mil for both nitrogen and carbon. In addition to carbon and nitrogen isotopes from the same sample, continuous flow will also report %C and %N data.

The lab runs one replicate per 10 samples, and any anomalous results are rerun. As a check on the combustion and cryogenic distillation steps, a laboratory standard is run every 15 samples. This standard is either peptone, a hydrolyzed animal protein from Sigma Chemical Company, glycine, or citrus leaves, SRM 1572. Both have been well documented by several stable isotope laboratories and their isotopic values are well known. Its value must be within 0.15 per mil of its documented value. If it does not, the samples preceding the standard are considered suspect and rerun.

The the addition of the GVI Instrumentation precision of water samples (oxygen-18 and deuterium) has improved significantly and is extremely good. The lab generally runs duplicates of all samples for oxygen-18 if there is enough water. The precision is usually 0.1 permil or better. An internal lab standard is run after every 4 client samples as a check on the instrumentation. Deuterium samples are run with the ChromeHD pyrolysis system. Three injections are done with each sample, with the first injection discarded, due to memory effects in the system. The standard is Boston University deionized water, collected in batch fashion and stored. The water standard is within 0.2 per mil of its long-term value for deuterium and 0.2 per mil for oxygen-18. In addition, the metabolic samples are inputted to a spreadsheet that calculates FMR (Field Metabolic Rates). If calculated values are not within acceptable ranges, the suspect samples are rerun. The calculations are based on the equations of Lifson and McClintock (1966), as modified by Kenneth Nagy, UCLA.

For carbonate samples, NBS-20 and Carera-Z are used as two point calibration standards. Precision is currently 0.05 permil for carbon-18 and 0.06 permil for oxygen-18. CO2 air and breath samples are calibrated using atmospheric air and a 1% CO2/helium mix gas. The mix gas was checked against calibration gases obtained from Oztech Corporation, Texas.

Data is presented in a tabular form and can be sent by fax, mail or email. The sheet includes sample ID, mass/volume used, isotopic value and % organics (if applicable). All isotopic data are rounded to 2 decimal places.

We request that a sample list be included with all samples and that all samples be clearly identified. This allows the Laboratory Manager to look over the data and compare the isotope values against generally accepted values for that type of sample. Any samples that appear anomalous are rerun if possible to check their values; if preloaded, they are flagged as anomalous for the client.